The AURA Ductless Fume Enclosure
consists of a head section with an Advanced Monitoring System, a variable
speed blower and fluorescent lighting. There are chemically-resistant
acrylic panels and a deep spill tray. Use two Hydrocarbon filters and two
Hydrocarbon Safety filters if you are using liquid chemicals in the
enclosure. If you are using only latent print powders, you need two HEPA
filters and one HEPA Safety filter. If you are going to use a mixture of
powders and liquid chemicals, you need two Hydrocarbon filters to screen the
fumes from the liquids and one HEPA Safety filter to screen the powders. It
is recommended that the pre-filter be changed about every 3-6 months,
depending on your usage. With the Hydrocarbon filter, safely use solvents,
such as Glacial Acetic Acid, Acetonitrile, Ethyl Alcohol, Methylene
Chloride, Petroleum Ether, Acetone, Cyclohexane, 2-Propanol, Ethyl Acetate
and Xylene among others.
Specifications:
- Dimensions: 42" w x 28" d x
48" h
- Weight without filters: 166
pounds
- Construction: Aluminum Head
Section and Frame coated with Polyurethane Paint.
- Fiberglass Spill Tray
- Adjustable Air Flow
- UL Approved Brushless and
Sparkless Centrifugal Motor Blower
- Average Face Velocity: 125
lfpm
- Average Face Velocity with
Lower Sash Open: 85 lfpm
- Internal pre-filter change
- 2-year Manufacturer's
Warranty
- Made in the USA
Advanced Monitor Specifications:
- Senses hydrocarbon and air
velocity levels
- Digital display of airflow
levels
- Displays filter type and
tracks the amount of time in use
- A sensor adjusts the blower
speeds based on the sash height
- Audible and visual alarm of
low airflow and hydrocarbon breakthrough
- Filter saturation alarms to
correspond to the solvents used and their exposure limit levels
Solvents only: order two
Hydrocarbon filters and two Hydrocarbon Safety filters
Powders only: order two HEPA filters and one HEPA Safety filter
Solvents and Powders: order two Hydrocarbon and one HEPA Safety filter |